| Customization: | Available |
|---|---|
| Applicable Departments: | Oral Department of Internal Medicine |
| Certification: | CE, FDA |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| Top working voltage | Focus value | Target angle | Fuse | Perspective loading | Anode thermal capacity | ||
| Current | Voltage | Working frequency | |||||
| 85KV | 1.5mm | 23° | 2.5 A | 2.8±0.7V | 50~60Hz | 150W | 9 KJ |
| X-ray Generator | |
| X-ray tube | XD2-1.4/85 |
| Focus value | 1.5mm |
| Total Al equivalent | 2mmAL |
| Tube voltage | 60KV |
| Tube current | 8mA |
| PROBLEM | JUDGEMENT | SOLUTION |
| Indicator doesn't light when power on | Check the power supply Check power cable If the core broken Check power switch cable Check if switch broken |
Be sure of power supply Change cable Change core Change switch cable Change switch |
| When press exposure button, it doesn't work | Check diode, buzzer and relay | Change them if broken |
| Controller and timer work but no X-ray output | Check cable Test tube head |
Change when broken |
| Machine can work but getting bad quality photo | Check the film Wrong exposure time Wrong developing method Low voltage |
Make sure the quality of film Take correct exposure time and developing time Use UPS to make sure the voltage 220V |









