PROBLEM | JUDGEMENT | SOLUTION |
Indicator doesn't light when power on | Check the power supply Check power cable If the core broken Check power switch cable Check if switch broken |
Be sure of power supply Change cable Change core Change switch cable Change switch |
When press exposure button, it doesn't work | Check diode, buzzer and relay | Change them if broken |
Controller and timer work but no X-ray output | Check cable Test tube head |
Change when broken |
Machine can work but getting bad quality photo | Check the film Wrong exposure time Wrong developing method Low voltage |
Make sure the quality of film Take correct exposure time and developing time Use UPS to make sure the voltage 220V |